Oct,02

ASME MFC-11M:2003 pdf download

ASME MFC-11M:2003 pdf download

ASME MFC-11M:2003 pdf download.MEASUREMENTOF FLUID FLOWBY MEANS OFCORIOLIS MASS FLOWMETERS
1 SCOPE
This Standard, ASME MFC-11M, gives guidelines for the selection, installation, calibration, and operation of Coriolis meters for the determination of mass flow, den- sity, volume flow, and other related parameters of flow- ing fluids. It also gives appropriate considerations re- garding the fluids to be measured. The content of this Standard is primarily applied to the metering of liquids. This Standard also gives guidance, within specified lim- its, to the metering of other fluids, mixtures of solids or gas inliquids, andmixtures ofliquids. AlthoughCoriolis meters may be used for gas measurement, specific guid- ance for gas measurement is not within the scope of this Standard.
2 TERMS AND DEFINITIONS
The terminology and symbols (Table 1) used in this Standard are in accordance with ASME MFC-1M. Terminology not defined in ASME MFC-1M but used in ASME MFC-11M are also defined in this paragraph. Some items from ASME MFC-1M are also listed in this paragraph for easier reference. accuracy ofmeasurement: the degree of freedom from er- ror; the degree of conformity of the indicated value to the true value of the measured quantity (see ASME MFC-1M). calibration: for the purpose of this Standard, calibration strictly refers to the procedure by which the flowmeter is checked against a traceable reference and does not include adjustment to the calibration factors. calibration factor(s): numerical factor(s), also called flow calibration factors, unique to each primary device and determined by flow calibration, which when pro- grammed into the transmitter, enables the meter to per- form to its stated specification (see the note in para. A1 of Appendix A for a definition of calibration). density calibration factor(s): calibration factor(s) asso- ciated with density measurement.Coriolis meter: a device consisting of a flow sensor (pri- mary device) and a transmitter (secondary device), which measures the mass flow by means of the Coriolis force generated by flowing fluid through oscillating tube(s); it may also provide measurements of density and temperature. cross-talk: if two or more Coriolis meters are to be mounted close together, interference through mechani- cal coupling may occur. This is often referred to as cross- talk. The manufacturer should be consulted for methods of avoiding cross-talk. elemental error: the bias and/or precision error associated with a single source or process in a chain of sources or processes (see ASME MFC-1M). flashing: the formation ofvapor bubbles in a liquid when the line pressure falls to or below the vapor pressure of the liquid, often due to local lowering of pressure be- cause of an increase in the liquid velocity (see ASME MFC-1M). flow rate: the quantity of fluid flowing though a cross section of a pipe per unit of time (see ASME MFC-1M). mass flow rate (q m ): the rate of flow of fluid mass through a cross section of a pipe (see ASME MFC-1M). volume flow rate (q v ): the rate of flow of fluid volume through a cross section of a pipe (see ASME MFC-1M). flow sensor (primary device): a mechanical assembly con- sisting of an oscillating tube(s), drive system, measure- ment sensor(s), supporting structure, flanges/fittings, and housing. drive system: means for inducing the oscillation of the tube(s). housing: environmental protection of the flow sensor. measurement sensor: sensor to detect the Coriolis ef- fect and to measure the frequency of the tube oscilla- tions. oscillating tube(s): tube(s) through which the fluid to be measured flows.transmitter (secondary device): electronic system provid- ing the drive and transforming the signals from the flow sensor to give output(s) of measured and inferred parameters; it also provides corrections derived from parameters such as temperature. uncertainty of measurement: the range within which the true value of the measured quantity can be expected to lie with a suitable highprobability (see ASME MFC-1M). zero offset: flow measurement indicated under zero flow conditions. zero stability: maximumexpected magnitude ofthe meter output at zero flow after the zero adjustment procedure has been completed, expressed by the manufacturer as an absolute value in mass per unit time.

Download
The previous

ASME Y14.1:2012 pdf download

The next

ASME Y14.6:2001 pdf download

Related Standards